Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
Avaliação do tempo de atraso do índice bispectral na resposta à indução da anestesia: estudo observacional

Rev. bras. anestesiol; 69 (4), 2019
Publication year: 2019

Abstract Background and objectives According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results The difference between the predicted and the real BIS was in average 30.09 ± 18.73 s. Conclusion Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.
Resumo Justificativa e objetivos De acordo com o fabricante, o índice bispectral (BIS) tem um tempo de processamento de cinco a dez segundos. Estudos que avaliaram esse tempo de processamento sugeriram atrasos mais longos. Nós avaliamos o tempo de atraso na resposta do BIS. Métodos Com base em dados clínicos de 45 pacientes, calculamos a diferença entre o tempo de atraso previsto e real do índice bispectral durante um período fixo de três minutos após o momento em que o BIS caiu abaixo de 80 durante a indução da anestesia geral com propofol e remifentanil. Resultados A diferença entre o BIS previsto e real foi em média 30,09 ± 18,73 segundos. Conclusão Nossos resultados sugerem que o atraso no processamento do índice bispectral pode ser muito maior do que o declarado pelo fabricante, um fato com implicações clínicas.

More related