Evaluación de la susceptibilidad directa desde hemocultivos positivos utilizando el sistema Vitek 2: comparación de dos protocolos rápidos
Evaluation of antimicrobial susceptibility testing directly from positive blood cultures using the Vitek 2 system: comparison of two rapid protocols
Rev. chil. infectol; 34 (2), 2017
Publication year: 2017
Antimicrobial susceptibility testing using Vitek® 2 (bioMérieux) cards inoculated directly from positive blood cultures was compared with the standard method for gramnegative rods. Two different protocols for the inoculum preparation were evaluated. Protocol 2 was faster and easier to perform. Rapid susceptibility testing was accurate and reduced time to results in 24 h.
Se comparó el ensayo de susceptibilidad a los antimicrobianos utilizando las tarjetas Vitek® 2 (bioMérieux) inoculadas directamente de hemocultivos positivos con el método estándar para bacilos gramnegativos. Se evaluaron dos protocolos diferentes para la preparación del inóculo. El segundo protocolo fue más rápido y fácil de realizar. Las pruebas de susceptibilidad rápida fueron precisas y redujeron el tiempo de los resultados en 24 h.