A rare case of kinked reinforced endotracheal tube in an intensive care unit: a case report
Un caso infrecuente de acodamiento de un tubo endotraqueal reforzado en una unidad de cuidados intensivos: reporte de un caso

Rev. colomb. anestesiol; 48 (1), 2020
Publication year: 2020

Abstract Reinforced endotracheal tubes (ETTs) are regularly applied in anesthetic practices to prevent compression or kinking. Although these tubes are designed to bend easily and prevent obstruction, they still carry a potential hazard of being obstructed following external forces. In this article, we reported an unusual case in which a reinforced ETT was collapsed due to the patient bite. The patient's bite force on the tube resulted in obstruction, hypoxia, and desaturation. To overcome this near-fatally condition, we removed the blocked reinforced ETT.
Resumen Los tubos endotraqueales reforzados (TET) se usan generalmente en la práctica anestésica para evitar la compresión o el acodamiento. Aun cuando dichos tubos están diseñados para flexionarse fácilmente y prevenir obstrucciones, de todos modos, existe la posibilidad de que se obstruyan, como consecuencia de fuerzas externas. En el presente trabajo reportamos un caso inusual en el cual se colapsó el TET a causa de la mordida del paciente. La fuerza de mordida sobre el tubo produjo obstrucción, hipoxia y desaturación. Con el fin de superar esta condición casi fatal, retiramos el TET reforzado bloqueado.