Accuracy of resonance frequency analysis by third generation Osstell®
Precisión del análisis de frecuencia de resonancia mediante Osstell® de tercera generación
Int. j. odontostomatol. (Print); 9 (3), 2015
Publication year: 2015
The Resonance Frequency Analysis by Osstell® is used in the assessment of primary and secondary implant stability. The aim of this study was to determine its accuracy as a method of testing, checking its repeatability and reproducibility. A pilot study was performed whereby implants were placed in a cow rib to compare several ISQ measurements with the same SmartPeg transducer (repeatability), or with different SmartPegs (reproducibility). Statistical analysis of the data showed a high correlation of the ISQ values obtained in both assays, which suggests that Osstell®analysis can provide high repeatability and reproducibility.
El análisis de frecuencia de resonancia (AFR) por Osstell® se utiliza en la evaluación de la estabilidad primaria y secundaria del implante. El objetivo del presente estudio fue establecer la precisión del AFR mediante Osstell©, como método de prueba empleado en la determinación de la estabilidad implantaria. Se realizó un estudio experimental in vitro, en el que se colocaron implantes en una costilla de vaca y compararon diferentes mediciones del coeficiente de estabilidad del implante (ISQ) obtenidas sobre estos implantes con un mismo transductor SmartPeg (repetibilidad) o con diferentes SmartPegs (reproductibilidad). El análisis estadístico de los datos mostró una elevada correlación de los ISQ obtenidos en ambos ensayos, lo cual sugiere una reproductibilidad y repetibilidad elevada del análisis mediante Osstell.