Perfuração de artéria coronária com rotablator
Coronary artery perforation with rotational coronary atherectomy

Arq. bras. cardiol; 66 (6), 1996
Publication year: 1996

The authors describe a rare case of circumflex coronary artery perforation during rotational coronary atherectomy complicated with cardiac tamponade and good outcome. The possible causes of perforation are discussed and the burr oversize (burr/artery ratio was 0.58) was refused. Shortening and artery plicature (accordeon effect) might have been the cause of this event. Quantitative measurement was made in order to strengthen this hypothesis. It is emphasized the importance of selecting lesions that should be submitted to rotational coronary atherectomy
É relatado caso raro de perfuração da artéria circunflexa durante aterectomia rotacional (AR), complicado com tamponamento cardíaco, que evoluiu favoravelmente. São discutidas as prováveis causas, afastando a possibilidade da ocorrência de um superdimensionamento da oliva utilizada (relação de diâmetro oliva-artéria foi de 0,58), e é sugerido o encurtamento e plicatura da artéria (efeito sanfona) como provável causa dessa perfuração. São feitas medidas quantitativas para reforçar essa hipótese, e ressaltam-se os cuidados necessários na escolha das lesões a serem submetidas à ablação com AR

More related