Precisão e exatidão das dosagens dos lípides sangüíneos em equipamento portátil (Cholestech-LDX)
Precision and accuracy of blood lipid analyses by a portable device (Cholestech-LDX)

Arq. bras. cardiol; 66 (6), 1996
Publication year: 1996

PURPOSE:

To verify whether precision and accuracy of lipids analyses by a new portable device, Cholestech-lipid desktop analyzer (LDX), were in agreement with the guidelines of the National Cholesterol Education Program (NCEP).

METHODS:

Serum samples from 45 outpatients were collected for the determination of total Cholesterol (TC), HDL-cholesterol (HDL-C) and triglycerides (TG). These samples were analysed simultaneously by the Cholestech-LDX, and by the automatic enzymatic methods routinely used at the Heart Institute's laboratory. Precision was determined by repeating 20 times the evaluation of the same sample of venous blood. Accuracy was established confronting the values of the lipids variables obtained with Cholestech-LDX against the values determined by the automatic enzymatic routine.

RESULTS:

Accuracy for TC was 1.60 (NCEP < or = 3), for HDL-C was -2.74 (NCEP < or = 6) and for TG was 2.11 (NCEP < or = 5). Precision for CT was 3.05 (NCEP < or = 3), for HDL-C was 1.05 (NCEP < or = 6) and for TG was 2.65 (NCEP < 5).

CONCLUSION:

Precision and accuracy of lipids evaluation by the Cholestech-LDX are within the guidelines of the National Cholesterol Education Program. Therefore the cholestech-LDX seems to be a reliable alternative to the conventional biochemical routine, allowing population screenings.
Objetivo - Verificar se a precisão e exatidão das dosagens de lípides, realizadas em equipamento portátil, Cholestech-lipid desktop analyser (LDX), concordam com os critérios estabelecidos pelo National Cholesterol Education Program (NCEP). Métodos - Amostras de soro de 45 pacientes ambulatoriais, coletadas para dosagem do colesterol total (CT), HDL-colesterol (HDL-C) e triglicérides (TG), foram analisadas simultaneamente pelo Cholestech-LDX e por métodos enzimáticos automatizados utilizados no Laboratório Clínico do Instituto do Coração. A precisão foi determinada repetindo 20 dosagens da mesma amostra de sangue venoso. A exatidão foi estabelecida confrontando os valores das variáveis lipídicas obtidos no CholestechLDX com aqueles determinados pela rotina laboratorial automatizada. Resultados - A precisão e exatidão das dosagens das variáveis lipídicas obtidas pelo Cholestech-LDX concordam com as recomendações do NCEP. Exatidão obtida para o CT foi 1,60% (NCEP <3%), para o HDL-C foi 2,74% (NCEP <6%) e para o TG foi 2,11% (NCEP <5%). Precisão obtida para o CT foi 3,05% (NCEP <3%), para o HDL-C foi 1,05 (NCEP <6%) e para o TG foi 2,65% (NCEP <5%). Conclusão - Precisão e exatidão das dosagens de lípides sangüíneos pelo Cholestech-LDX concordam com as recomendações do NCEP, tornando-se alternativa confiável à rotina laboratorial convencional, permitindo avaliação de amostras populacionais