Rev. bras. cir. cardiovasc; 32 (6), 2017
Publication year: 2017
Abstract Introduction:
Tissue adhesives can be used as adjacent to sutures to drop or avoid bleeding in cardiovascular operations. Objective:
To verify the efficiency of fibrin and cyanoacrylate adhesive to seal arterial sutures and if the adhesives penetrate through suture line to the inner of arteries. Methods:
20 abdominal aorta segments of pigs were divided into two groups according to the adhesive which would be used as adjacent to the suture. In every arterial segment an arteriotomy was done, followed by a conventional artery closure. Afterwards a colloidal fluid was injected inside the arterial segment with a simultaneous intravascular pressure monitoring up to a fluid leakage through the suture. This procedure was repeated after application of one of the adhesives on the suture in order to check if the bursting pressure increases. The inner aorta segments also were analyzed in order to check if there was intraluminal adhesive penetration. Results:
In Suture 1 group, the mean arterial pressure sustained by the arterial suture reached 86±5.35 mmHg and after the fibrin adhesive application reached 104±11.96 (P<0.002). In the Suture 2 group, the mean arterial pressure sustained by the suture reached 83±2.67 mmHg and after the cyanoacrylate adhesive application reached 152±14.58 mmHg (P<0.002). Intraluminal adhesive penetration has not been noticed. Conclusion:
There was a significant rise in the bursting pressure when tissue adhesives were used as adjacent to arterial suture, and this rise was higher if the cyanoacrylate adhesive was used. In addition, the adhesives do not penetrate through the suture line into the arteries.