Effects of different torque levels on the implant-abutment interface in a conical internal connection
Braz. oral res. (Online); 30 (1), 2016
Publication year: 2016
Abstract The fit of the implant-abutment interface was assessed by the metallographic technique and by scanning electron microscopy (SEM), using solid abutment types at different torque levels. Forty Morse taper connections and forty solid abutments were used at different torque levels (repeated after 10 minutes) in the following groups (n = 10): 25 Ncm (group g1), 30 Ncm (group g2), 35 Ncm (group g3), and 40 Ncm (group g4). The samples were embedded in a metallographic resin, sectioned lengthwise, and polished. SEM images were used to measure the linear contacts and the fits between abutments and the internal walls of the implant.