Eventos adversos associados ao uso de cateteres arteriais em unidades de terapia intensiva: um estudo transversal
Adverse events associated with the use of arterial catheters in intensive care units: a cross-sectional study

Publication year: 2024
Theses and dissertations in Portugués presented to the Universidade Federal do Rio de Janeiro. Centro de Ciências da Saúde. Escola de Enfermagem Anna Nery. Programa de Pós-Graduação em Enfermagem, to obtain the academic title of Mestre. Leader: Paes, Graciele Oroski

Introdução:

A monitorização de pressão arterial invasiva e a manipulação do circuito são processos fundamentais na assistência de enfermagem em terapia intensiva, mas práticas inadequadas podem gerar riscos significativos para o paciente, incluindo eventos adversos.

Objetivos:

Geral: Analisar os fatores contribuintes para a ocorrência de eventos adversos associados ao uso de cateteres arteriais, à luz da segurança do paciente.

Específicos:

1 - Verificar a ocorrência de eventos associados ao uso de cateteres arteriais em terapia intensiva. 2 - Identificar os fatores contribuintes para a ocorrência de eventos adversos associados ao uso de cateteres arteriais em terapia intensiva; 3 - Correlacionar a ocorrência de eventos adversos associados ao uso de cateteres arteriais com os fatores contribuintes relacionados à segurança do paciente; 4 - Discutir a implicação dos fatores contribuintes para a prática segura de enfermagem, à luz da segurança do paciente.

Método:

estudo transversal, com dados retrospectivos, de caráter analítico e de natureza quantitativa. A amostra documental foi composta por todas as notificações de eventos adversos relacionados a cateteres arteriais entre dezembro de 2019 e janeiro de 2023 nas terapias intensivas da instituição coparticipante. Para a complementação de dados clínicos dos pacientes, foram utilizados os prontuários eletrônicos e físicos do hospital. Realizada análise descritiva e inferencial com cálculos de frequências e testes de associação, como qui-quadrado e exato de Fisher. Em toda análise foi considerado o nível de significância máximo de 10% (p-valor < 0,1). O estudo seguiu as diretrizes do Checklist STROBE.

Resultados:

O estudo incluiu 39 notificações de eventos adversos com cateteres arteriais, com uma taxa de ocorrência de 1,25%. O incidente de maior prevalência foi a perda inadvertida (41%), seguido por formação pseudoaneurisma e/ou hematoma (25,6%), e obstrução (20,51%). Quanto às consequências dos eventos, a necessidade de nova punção ocorreu em 61,5% dos casos, sendo a maioria dos danos classificados como leves (53,8%). Recursos humanos (p-valor 0,069), necessidade de sedação (p-valor 0,011) e sítio de inserção (p-valor 0,074) foram fatores que estiveram associados à perda acidental do dispositivo. Fatores relacionados ao paciente (p-valor 0,079) e ao método de punção (p-valor 0,079) estiveram relacionados à formação de pseudoaneurisma e hematomas. E recursos humanos (p-valor 0,001), utilização de aminas vasoativas (p-valor 0,032) e o sítio de inserção (p-valor 0,012) estiveram associados à obstrução do sistema.

Conclusão:

A partir dos resultados foi possível elucidar potenciais fragilidades que podem comprometer a segurança do paciente crítico em uso de monitorização de pressão arterial, como método de punção, manejo do circuito e características clínicas dos pacientes. Fatores como recursos humanos, uso de sedação e o local de inserção do cateter estiveram vinculados à perda acidental do mesmo. Além disso, características relativas ao paciente e ao método de punção foram associadas à formação de pseudoaneurisma e/ou hematoma. Por fim, a combinação de fatores como recursos humanos, uso de aminas vasoativas, método de manuseio e o local de inserção demonstrou associação com a obstrução do cateter. Futuros estudos de acompanhamento com amostras maiores podem fortalecer a compreensão do fenômeno.

Introduction:

Invasive blood pressure monitoring and circuit manipulation are fundamental processes in intensive care nursing, but inadequate practices can generate significant risks for the patient, including adverse events.

Objectives:

General: To analyze the contributing factors to the occurrence of adverse events associated with the use of arterial catheters, in the light of patient safety.

Specific:

1 - To verify the occurrence of events associated with the use of arterial catheters in intensive care; 2 - To identify the contributing factors for the occurrence of adverse events associated with the use of arterial catheters in intensive care; 3 - To correlate the occurrence of adverse events associated with the use of arterial catheters with the contributing factors related to patient safety; 4 - To discuss the implication of the contributing factors for safe nursing practice, in the light of patient safety.

Method:

cross-sectional study, with retrospective data, of an analytical and quantitative nature. The documentary sample consisted of all reports of adverse events related to arterial catheters between December 2019 and January 2023 in the intensive care units of the co-participating institution. The hospital's electronic and physical medical records were used to supplement the patients' clinical data. Descriptive and inferential analysis was carried out using frequency calculations and association tests such as chi-square and Fisher's exact test. A maximum significance level of 10% (p-value < 0.1) was considered in all analyses. The study followed the guidelines of the STROBE Checklist.

Results:

The study included 39 reports of adverse events with arterial catheters, with an occurrence rate of 1.25%. The most prevalent incident was inadvertent loss (41%), followed by pseudoaneurysm and/or hematoma formation (25.6%), and obstruction (20.51%). As for the consequences of the events, the need for a new puncture occurred in 61.5% of cases, with most of the damage being classified as mild (53.8%). Human resources (p-value 0.069), need for sedation (p-value 0.011) and insertion site (p-value 0.074) were factors associated with accidental device loss. Factors related to the patient (p-value 0.079) and the puncture method (p-value 0.079) were related to the formation of pseudoaneurysms and hematomas. Human resources (p-value 0.001), the use of vasoactive amines (p-value 0.032) and the insertion site (p-value 0.012) were associated with system obstruction.

Conclusion:

From the results, it was possible to elucidate potential weaknesses that could compromise the safety of critically ill patients using blood pressure monitoring, such as the method of puncture, circuit management and the clinical characteristics of the patients. Factors such as human resources, the use of sedation and the catheter insertion site were linked to accidental catheter loss. In addition, characteristics relating to the patient and the method of puncture were associated with the formation of pseudoaneurysms and/or hematomas. Finally, the combination of factors such as human resources, use of vasoactive amines, handling method and insertion site was associated with catheter obstruction. Future follow-up studies with larger samples could strengthen our understanding of this phenomenon.

Introducción:

La monitorización invasiva de la presión arterial y la manipulación del circuito son procesos fundamentales en la enfermería de cuidados intensivos, prácticas inadecuadas pueden generar riesgos significativos para el paciente, incluyendo eventos adversos.

Objetivos:

General: Analizar los factores que contribuyen a la ocurrencia de eventos adversos asociados al uso de catéteres arteriales, a la luz de la seguridad del paciente.

Específicos:

1 - Verificar la ocurrencia de eventos asociados al uso de catéteres arteriales en cuidados intensivos; 2 - Identificar los factores contribuyentes para la ocurrencia de eventos adversos asociados al uso de catéteres arteriales en cuidados intensivos; 3 - Correlacionar la ocurrencia de eventos adversos asociados al uso de catéteres arteriales con los factores contribuyentes relacionados a la seguridad del paciente; 4 - Discutir la implicación de los factores contribuyentes para la práctica segura de enfermería, a la luz de la seguridad del paciente.

Método:

Estudio transversal, con datos retrospectivos, de carácter analítico y cuantitativo. La muestra documental consistió en todos los informes de eventos adversos relacionados con catéteres arteriales entre diciembre de 2019 y enero de 2023 en las unidades de cuidados intensivos de la institución coparticipante. Se utilizaron las historias clínicas electrónicas y físicas del hospital para complementar los datos clínicos de los pacientes. Se realizaron análisis descriptivos e inferenciales mediante cálculos de frecuencias y pruebas de asociación como chi-cuadrado y la prueba exacta de Fisher. Consideró un nivel de significación máximo del 10% (valor p < 0,1). El estudio siguió las directrices de la STROBE Checklist.

Resultados:

El estudio incluyó 39 notificaciones de acontecimientos adversos con catéteres arteriales, con una tasa de ocurrencia del 1,25%. El incidente más prevalente fue la pérdida inadvertida (41%), seguido de la formación de pseudoaneurismas y/o hematomas (25,6%) y la obstrucción (20,51%). En cuanto a las consecuencias de los sucesos, la necesidad de una nueva punción se produjo en el 61,5% de los casos, clasificándose la mayoría de los daños como leves (53,8%). Los recursos humanos (p-valor 0,069), la necesidad de sedación (p-valor 0,011) y el lugar de inserción (p-valor 0,074) fueron factores asociados a la pérdida accidental del dispositivo. Los factores relacionados con el paciente (p-valor 0,079) y el método de punción (p-valor 0,079) se relacionaron con la formación de pseudoaneurismas y hematomas. Los recursos humanos (p-valor 0,001), el uso de aminas vasoactivas (p-valor 0,032) y el lugar de inserción (p-valor 0,012) se relacionaron con la obstrucción del sistema.

Conclusiones:

A partir de los resultados, fue posible dilucidar las debilidades que podrían comprometer la seguridad de los pacientes críticos que utilizan la monitorización de la presión arterial, como el método de punción, el manejo del circuito y las características clínicas de los pacientes. Factores como los recursos humanos, el uso de sedación y el lugar de inserción del catéter se relacionaron con la pérdida accidental del catéter. Además, las características relacionadas con el paciente y el método de punción se asociaron a la formación de pseudoaneurismas y/o hematomas. Por último, la combinación de factores como los recursos humanos, el uso de aminas vasoactivas, el método de manipulación y el lugar de inserción se asoció a la obstrucción del catéter. Futuros estudios de seguimiento con muestras más amplias podrían reforzar la comprensión del fenómeno.